Main application scenarios: human body impedance analysis, AC lipid measurement.
CS1256 is a digital filter that includes an ADC signal chain, a body impedance measurement module (BIM), a Sigma delta ADC, and a digital filter. The ADC signal link includes input MUXP/MUXN and programmable low noise gain amplifier (PGA). The MUXP/MUXN has three internal input channels. PGA and ADC have multiple gain options, and digital filters can be configured for multiple output rates. The human body impedance measurement module uses a sinusoidal excitation source to convert the human body impedance into a voltage signal and transmits it to the ADC signal chain for measurement. It supports multi-electrode and multi-frequency human body impedance measurement.
BIM
-Supports 4-electrode measurement
-Supports 5 kHz/10 kHz/25 kHz/50 kHz/100 kHz/250 kHz multi gear frequency measurement
-Supports absolute impedance value and phase angle measurement
ADC
-24-bit resolution
-Output speed: 10~1280 Hz, 8-gear optional
19.5-bit effective at 2.35 V reference, 40 Hz rate and 128x gain
LDO and internal reference voltage
-Self-contained LDO, output 2.35/2.45/2.8/3.0 V selectable, accuracy ±1%
-Self-contained low drift reference, internal reference voltage 2.048 V selectable, accuracy ±1%
Supports performance, normal, low power, and sleep modes
Supports voltage measurement, temperature measurement, BIM measurement, and manual measurement modes with single command switching
Low drift on-chip clock
Three-wire serial communication
Core sea 8-bit MCU/SOC integrated development environment
1. Integrated C compiler, assembler, linker, debugger
2. Support standard C language and assembly language development 3 Example code of integrated chips